Nice to meet you.<br>I really appreciate your help.<br>Our calibration failed for three reasons.<br>I use pH4/7 acid calibration as an example.<br>1. Measure the value in buffer 1(pH7). If the value is not read, the calibration will be judged as a failure, and the result will be Buf 1 0. 00 Buf 2 0。 00 slope 0. 00.<br>2. Measure the value in buffer 2(pH4). If no value is read, the calibration will be judged as failure, and the result will be Buf 1-14. 05 (this is the assumption of the measured value) Buf 2 0. 00 slope 0. 00.<br>3. After measuring two calibration points, the notch is not in 52-62, and we will also determine that the calibration failed.<br><br>In addition, we would like to discuss the first and second points to see if there is any way to see why the same electrode passed and failed in a short time from the log of CM442R.<br>All the calibration parameters are the same. We just recalibrate the automatic calibrator. The service industry includes store services and administrative services.