Starting from the Initialize the hard ware.
Idle condition, wait for user commands through keys keys or mouse
Model Selection when users select the version of the meter to be calibrated. Program test will show the picture of the meter that class, and return to the state Idle.
.When the user press work Load File program will load the specific data of test to the table and return to the state Idle.
when users press the button. Run. Program will check the characters available in the test channel Serial number with name of the calibration. That was to identify whether or not.Unspecified will show a warning and go back to the Idle key again.
Check end step if not yet Show range selection Send command,, To wave TEK, Set lower & upper spec Wait very Actual back to a condition Idle. But if done back at the state Idle.
.When a user key readings from the meter after calibration, hit OK Comparison Actual value Vs Accept value. In, tolerance. Accept. Write data to table. And Save data from table report file. Go Check end step. But if Reject Wait confirm. If Yes. Write data to table. And Save data from table report file.If No Clear value Wait very Actual back to a condition Idle.
order ON / OFF use Send command ON / OFF output WAVETEK 9105
.