Operate the blood leak detector in the standby mode after completing the cleaning process.
After operating the detector for 5 min or longer in the standby process, turn the adjustment
volume 1 of the blood leak detector to the left. Adjust the REAL blood leak value in the range of 30
to 40 ppm.
Note 1: Make sure that no air bubbles are inside the cartridge of the blood leak detector.
Note 2: When the adjustment volume 1 is turned excessively, the displayed value is slightly
lower than the actual value.